7SI DIN Compact Digital Indicator


Supply Voltage: 100-240 Vac (+10%, -15%), 50/60 Hz
or 24 Vac/Vdc (±10%)
Operating Ambient: 0-50°C, 20-85% RH non-condensing
Inputs: T/C Types B, C, D, E, G, L, J, K, N,
Platinel II, R, S, T and U (°C, °F);
Pt 100 3W RTD (°C, °F) Ranges:
mAdc, mVdc, Vdc
Logic Input: for Alarm Acknowledgement, requires contact rated at 0.5 mA, 5 Vdc minimum
Output Ratings: Output 1: Relay, 3A/250V, SPDT, Resistive
Direct or Reverse operation
Output 2: (Optional) Relay, 2A/250V, SPST, NO, Resistive
Direct or Reverse operation
Serial Comms: EIA RS-485 Modbus®, JBUS
Mounting: Panel Mount

Thermocouple and RTD Input Range Table
Input Type °C Range °F Range
TC B 0 to 1820 32 to 3300
TC C 0 to 2300 0 to 4170
TC D 0 to 2300 0 to 4170
TC E -100 to 800 -150 to 1470
TC G 0 to 2300 0 to 1470
TC L -100 to 900 -150 to 1650
TC J -100 to 1000 -150 to 1830
TC K -100 to 1370 -150 to 2500
TC N -100 to 1400 -150 to 2550
TC Platinel II -100 to 1400 -150 to 2550
TC R -50 to 1760 -60 to 3200
TC S -50 to 1760 -60 to 3200
TC T -200 to 400 -330 to 750
TC U -200 to 600 -330 to 1110

Alarm Standby Function
The 7SI can be programmed with an alarm masking function (Alarm standby), which puts the alarm in a standby condition during instrument powerup, when process low alarms are subject to false activation. Once the low alarm threshold is traversed for the first time the instrument reverts to standard operation.

Data Memory Function
The 7SI also has a “data memory” function which is enabled at powerup. It is able to store in memory the minimum and maximum measured value, which, by pushbutton or serial link, can be displayed on the front panel. This function can also be reset by pushbutton or link, deleting the old values and starting a new memorization period.