| Technical Specification Memocal 2000 |
|
| Case |
|
ABS, color similar to RAL 6038 |
| Self extinguishing degree |
V-0 according to UL |
| Input protection |
|
All measured and generated ranges are protected against fault connection to signals up to 30 V AC/DC |
| Terminals |
|
3 external screw FEMALE plugs dia. 4mm |
| Weight |
|
600g max. (1.4lb) |
| Power supply |
|
4 batteries AA size. Alkaline 1.5V or Ni-Cd 1.2 V or NI-MH 1.2V. Switching type AC adapter (100 to 240 V AC) |
| Battery life |
|
24 hours (average) with Ni-MH batteries |
| Recharging time |
|
12 hours |
| Insulation resistance |
|
>100MΩ |
| Isolation voltage |
|
1500 V r.m.s. |
| Common mode rejection ratio |
120dB@50/60Hz |
| Normal mode rejection ratio |
60dB@50/60Hz |
| Safety standard |
|
Meets EN60950/A2 |
| EMC standard |
|
Meets generic emissions standard EN50081-2 and immunity standard EN50082-2 for industrial environments |
| A/D conversion |
|
Dual slope integration |
| Sampling time |
|
500ms |
| Display update time |
|
500ms |
| Temperature drift |
|
0.0028%/°C or 28ppm/°C |
| Operating temperature |
|
0 to +40°C |
| Storage temperature |
|
-10 to +60°C |
| Humidity |
|
20% to 85% RH non condensing |
|
| Measuring
Inputs |
| mA and V inputs* |
|
|
| Standard ranges table |
|
| Range |
Resolution |
Ref. Accur. (±) |
-20mV to 20mV -200mV to 200mV -2V to 2V -20V to 20V |
1μV 10μV 100μV 1mV |
0.015% 0.015% 0.015% 0.020% |
-20mA to 20mA -130mA to 130mA |
1μA 10μA |
0.015% 0.020% |
|
| Range selection |
|
Automatic or manual |
| Input impedance |
|
10Ω for mA input >500 kΩ for 20V input |
| Square root extraction |
|
Programmable |
| Read-out |
|
Keyboard programmable from -20000 to 20000 |
| Decimal point |
|
Programmable in any position |
|
| RTD input |
| RTD type |
|
*PT100 3-wire connection / Ni 100 3-wire connection |
| Calibration |
|
Programmable according to IPTS-68 or ITS-90 |
| Line resistance |
|
Up to 20Ω/wire with no measurable error |
| Engineering unit |
|
°C or °F keyboard programmable |
| Measuring current |
|
100μA |
| Burn out |
|
Detection of the sensor open circuit and one or more wires open circuit |
| Standard range table for RTD PT100 |
| Range PT100 |
Resolution |
Ref. Accur. (±) |
-200°C to 850°C -328°F to 512°F 513°F to 1562°F |
0.1°C 0.1°F <0.2°F |
0.028% 0.012% 0.029% |
|
| |
| Standard range table for Ni 100 |
| Range Ni100 |
Resolution |
Ref. Accur. (±) |
-60°C to 350°C -76°F to 662°F |
0.1°C 0.1°F |
0.029% 0.029% |
|
| |
|
| |
|
Available only when the IPTS-68
standard has been selected |
|
|
| Ohms input |
| Standard range table |
| Range |
Resolution |
Ref. Accur. (±) |
| 0 to 800Ω |
0.1Ω |
0.025% |
|
|
| TX Measurement |
| The TX measurement is mA measurement with a 24V power supply generated by the instrument
and is used to calibrate 2, 3 or 4-wire transmitters. |
| Power supply |
|
24V DC (maximum current 24 mA) |
| Resolution |
|
1 μA |
| Reference accuracy |
|
0.015% |
| Input impedance |
|
10Ω for mA input |
| Input range |
|
0 to 20mA |
| Square root extraction |
|
Programmable |
| Read-out |
|
Keyboard programmable from -20000 to 20000 |
| Decimal point |
|
Programmable in any position |
| Burn out |
|
The instrument shows the ‘OPEN’ message when a burn out condition is detected |
|
| Thermocouples |
| Type |
|
B,E,J,K,L.N,Ni/Ni-18%Mo, PLII, R, S, T*, U, W, W3 and W5 keyboard programmable. |
| Engineering unit |
|
°C or °F keyboard programmable |
| Burn out |
|
Detection of the open input circuit (wires or sensor) with “OPEN” indication |
| Cold junction |
|
Automatic compensation |
| Cold junction compensation error |
±0.3 °C±0.005 °C/°C |
| External cold junction compensation |
Programmable value from -20 to 80°C |
| Input impedance |
|
>10MΩ |
| Calibration |
|
Programmable according to IPTS-68 or ITS-90 |
|
| ----Measuring Outputs---- |
| mA and V outputs |
|
|
| Standard range table |
|
| Range |
Resolution |
Ref. Accur. (±) |
-4mV to 20mV -40mV to 200mV -400mV to 2000mV -4V to 20V 0mA to 21mA |
1μV 10μV 100μV 1mV 1μA |
0.015% 0.015% 0.015% 0.020% 0.015% |
|
| Output impedance outputs |
50Ω for mV output / 0.5Ω for the other V |
| Maximum load for mA output |
500Ω |
| Range selection |
|
Automatic or manual |
| Quadratic signal generation |
Programmable |
| Read-out |
|
Keyboard programmable from -4000 to 20000 |
| Decimal point |
|
Programmable in any position |
|
| RTD simulation |
| RTD type |
|
PT100 3 wire connection / Ni 100 3 wire connection |
| Calibration |
|
Programmable according to IPTS-68 or ITS-90 |
| Engineering unit |
|
°C or °F keyboard programmable |
| |
|
|
| Standard range table for RTD PT100 |
|
| Range (PT100) |
Resolution |
Ref. Accur. (±) |
-200°C to 850°C -328°F to 512°F 513°F to 1562°F |
0.1°C 0.1°F <0.2°F |
0.029% 0.022% 0.025% |
|
| |
|
|
| Standard range table for Ni100 |
|
| Range (Ni100) |
Resolution |
Ref. Accur. (±) |
-60°C to 350°C -76°F to 680°F |
0.1°C 0.1°F |
0.036% 0.036% |
|
| |
|
|
| |
|
Available only when the IPTS-68 standard has been selected |
|
|
| Ohms simulation |
| Standard range table |
|
|
| Range |
Resolution |
Ref. Accur. (±) |
| 15 to 500Ω |
0.1Ω |
0.031% |
|
|
| Thermocouples |
| Type |
|
B,E,J,K,L.N,Ni/Ni-18%Mo, PLII, R, S, T, U, W, W3 and W5 keyboard programmable. |
| Engineering unit |
|
°C or °F keyboard programmable |
| Output impedance |
|
50Ω |
| Calibration |
|
See table |
| Standard ranges |
|
For ranges, accuracies and resolutions see thermocouples standard ranges table |
|
| Special Features |
| Backlight |
|
LED backlighting with manual ON and automatic (30 seconds) shut OFF |
| Self diagnostic |
|
At instrument start up |
| Peak detection |
|
Minimum and maximum Peak Picker |
|
| Sequential Function Routine |
This instrument is provided with 50 steps which can be
utilized to make up one or more programs
Each program can consist of a number of soak and ramp simulations measurements and standby steps. |
| Low battery test |
|
|
| A test is provided at 3 different levels: |
|
- Level 1: the instrument is fully operative
- Level 2: the instrument shows the message
“BATTERY LOW” on the upper display while
the lower display shows the measured or
generated value.
The instrument is still fully operative.
- Level 3: the instrument turns OFF
automatically and utilizes the remaining
power for memory back up only.
|
|
| Serial
Interface (optional) |
| Types |
Built-in: |
|
RS-232C |
| |
External: |
|
RS-232 to RS485 opto-isolated converter installed in the desk-top stand |
| Protocol type |
|
MODBUS, JBUS |
| Baud rate |
|
Keyboard programmable from 600 to 19200 BAUD |
| Byte format |
|
8 bit |
| Parity |
|
Even, odd or none programmable |
| Stop bit |
|
One |
| Address |
|
From 1 to 255 |
| Output voltage |
|
According to EIA standard levels |
|
| Logic
Inputs |
The MEMOCAL 2000 is equipped with 2 logic inputs.
They are mutually exclusive with serial interface.
These 2 logic inputs have the following functions: |
- Logic input 1: sequences to the next program step.
- Logic input 2: it allows to start or to suspend the program execution (RUN/WAIT)
|